User Contributed MET/CAL Procedure ============================================================================= INSTRUMENT: EMC 11 CW Sensitivity Sub DATE: 21-Dec-95 AUTHOR: User Contributed REVISION: ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 11 NUMBER OF LINES: 169 CONFIGURATION: HP 3325A ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 IEEE [TERM LF] 1.002 HEAD {} 1.003 HEAD {ÉÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍ»} 1.004 HEAD {º CW Sensitivity º} 1.005 HEAD {ÈÍÍÍÍÍÍÍÍÍÍÍÍÍÍÍͼ} 1.006 HEAD ÍÍ CW Sensitivity ÍÍ 1.007 JMP 2.001 1.008 EVAL 2.001 DISP Set the step attenuator for 110 dB of attenuation. 2.002 DISP Disconnect the 3325A from the attenuator. 2.003 IEEE [@3325]FU1;AM 1MR;FR 25KH 2.004 IEEE HPF:6000[13][10] 2.004 IEEE BW:0010[13][10] 2.004 IEEE ATT:0[13][10] 2.005 IEEE GTL[13][10] 2.006 IEEE [T0]CAL[13][10] 2.007 IEEE GTL[13][10] 2.008 IEEE [D5000][I][13][10][D1000][I][13][10] 2.009 MEME >1 2.010 DISP Connect the 3325A to the step attenuator. 2.011 MATH mem1 = m[1] + 3 2.012 DISP Adjust the attenuator for a reading of [mem1] 2.012 DISP on the UUT digital display. 2.013 MEMI Enter the attenuation setting from the step attenuator. 2.014 MATH mem = 60 - mem 2.015 MATH mem1 = -49 2.016 MEME 2.017 MEMC d +0U -100U BW=10H 3.001 DISP Disconnect the 3325A from the attenuator. 3.002 IEEE BW:0040[13][10] 3.003 IEEE GTL[13][10] 3.004 IEEE [T0]CAL[13][10] 3.005 IEEE GTL[13][10] 3.006 IEEE [D5000][I][13][10][D1000][I][13][10] 3.007 MEME >1 3.008 DISP Connect the 3325A to the step attenuator. 3.009 MATH mem1 = m[1] + 3 3.010 DISP Adjust the attenuator for a reading of [mem1] 3.010 DISP on the UUT digital display. 3.011 MEMI Enter the attenuation setting from the step attenuator. 3.012 MATH mem = 60 - mem 3.013 MATH mem1 = -43 3.014 MEME 3.015 MEMC d +0U -100U BW=40H 4.001 DISP Disconnect the 3325A from the attenuator. 4.002 IEEE BW:0100[13][10] 4.003 IEEE GTL[13][10] 4.004 IEEE [T0]CAL[13][10] 4.005 IEEE GTL[13][10] 4.006 IEEE [D5000][I][13][10][D1000][I][13][10] 4.007 MEME >1 4.008 DISP Connect the 3325A to the step attenuator. 4.009 MATH mem1 = m[1] + 3 4.010 DISP Adjust the attenuator for a reading of [mem1] 4.010 DISP on the UUT digital display. 4.011 MEMI Enter the attenuation setting from the step attenuator. 4.012 MATH mem = 60 - mem 4.013 MATH mem1 = -39 4.014 MEME 4.015 MEMC d +0U -100U BW=100H 5.001 DISP Disconnect the 3325A from the attenuator. 5.002 IEEE BW:0250[13][10] 5.003 IEEE GTL[13][10] 5.004 IEEE [T0]CAL[13][10] 5.005 IEEE GTL[13][10] 5.006 IEEE [D5000][I][13][10][D1000][I][13][10] 5.007 MEME >1 5.008 DISP Connect the 3325A to the step attenuator. 5.009 MATH mem1 = m[1] + 3 5.010 DISP Adjust the attenuator for a reading of [mem1] 5.010 DISP on the UUT digital display. 5.011 MEMI Enter the attenuation setting from the step attenuator. 5.012 MATH mem = 60 - mem 5.013 MATH mem1 = -35 5.014 MEME 5.015 MEMC d +0U -100U BW=200H 6.001 DISP Disconnect the 3325A from the attenuator. 6.002 IEEE BW:0500[13][10] 6.003 IEEE GTL[13][10] 6.004 IEEE [T0]CAL[13][10] 6.005 IEEE GTL[13][10] 6.006 IEEE [D5000][I][13][10][D1000][I][13][10] 6.007 MEME >1 6.008 DISP Connect the 3325A to the step attenuator. 6.009 MATH mem1 = m[1] + 3 6.010 DISP Adjust the attenuator for a reading of [mem1] 6.010 DISP on the UUT digital display. 6.011 MEMI Enter the attenuation setting from the step attenuator. 6.012 MATH mem = 60 - mem 6.013 MATH mem1 = -32 6.014 MEME 6.015 MEMC d +0U -100U BW=500H 7.001 DISP Disconnect the 3325A from the attenuator. 7.002 IEEE BW:1000[13][10] 7.003 IEEE GTL[13][10] 7.004 IEEE [T0]CAL[13][10] 7.005 IEEE GTL[13][10] 7.006 IEEE [D5000][I][13][10][D1000][I][13][10] 7.007 MEME >1 7.008 DISP Connect the 3325A to the step attenuator. 7.009 MATH mem1 = m[1] + 3 7.010 DISP Adjust the attenuator for a reading of [mem1] 7.010 DISP on the UUT digital display. 7.011 MEMI Enter the attenuation setting from the step attenuator. 7.012 MATH mem = 60 - mem 7.013 MATH mem1 = -29 7.014 MEME 7.015 MEMC d +0U -100U BW=1kH 8.001 DISP Disconnect the 3325A from the attenuator. 8.002 IEEE BW:2500[13][10] 8.003 IEEE GTL[13][10] 8.004 IEEE [T0]CAL[13][10] 8.005 IEEE GTL[13][10] 8.006 IEEE [D5000][I][13][10][D1000][I][13][10] 8.007 MEME >1 8.008 DISP Connect the 3325A to the step attenuator. 8.009 MATH mem1 = m[1] + 3 8.010 DISP Adjust the attenuator for a reading of [mem1] 8.010 DISP on the UUT digital display. 8.011 MEMI Enter the attenuation setting from the step attenuator. 8.012 MATH mem = 60 - mem 8.013 MATH mem1 = -24 8.014 MEME 8.015 MEMC d +0U -100U BW=2.5kH 9.001 DISP Disconnect the 3325A from the attenuator. 9.002 IEEE BW:5000[13][10] 9.003 IEEE GTL[13][10] 9.004 IEEE [T0]CAL[13][10] 9.005 IEEE GTL[13][10] 9.006 IEEE [D5000][I][13][10][D1000][I][13][10] 9.007 MEME >1 9.008 DISP Connect the 3325A to the step attenuator. 9.009 MATH mem1 = m[1] + 3 9.010 DISP Adjust the attenuator for a reading of [mem1] 9.010 DISP on the UUT digital display. 9.011 MEMI Enter the attenuation setting from the step attenuator. 9.012 MATH mem = 60 - mem 9.013 MATH mem1 = -21 9.014 MEME 9.015 MEMC d +0U -100U BW=5kH 10.001 DISP Disconnect the 3325A from the attenuator. 10.002 IEEE BW:WB[13][10] 10.003 IEEE GTL[13][10] 10.004 IEEE [T0]CAL[13][10] 10.005 IEEE GTL[13][10] 10.006 IEEE [D5000][I][13][10][D1000][I][13][10] 10.007 MEME >1 10.008 DISP Connect the 3325A to the step attenuator. 10.009 MATH mem1 = m[1] + 3 10.010 DISP Adjust the attenuator for a reading of [mem1] 10.010 DISP on the UUT digital display. 10.011 MEMI Enter the attenuation setting from the step attenuator. 10.012 MATH mem = 60 - mem 10.013 MATH mem1 = -11 10.014 MEME 10.015 MEMC d +0U -100U BW=WB 11.001 END